ESP-32HD-DTC, Miniature Analog Pressure Scanner




- Differential or Vented Gauge
- Dry Noncorrosive Gases
- Proof Pressure 4 Times Rated Pressure (≤ 30 psi)
- Proof Pressure 1.5 Times Rated Pressure (> 30 psi)
- 0.040" or 0.063" OD Stainless Steel Tubulation
- Digital Temperature Compensation (Option)
ESP Pressure Scanners are miniature electronic differential pressure measurement units consisting of an array of silicon piezoresistive pressure sensors, one for each pressure port. The sensors are mounted on a common hybrid glass substrate using a proprietary technique that maximizes long-term stability. The outputs of the sensors are electronically multiplexed through a single onboard instrumentation amplifier at rates up to 70,000 Hz using binary addressing. The multiplexed amplified analog output is capable of driving long lengths of cable to a remote A/D converter.
The ESP scanners also incorporate a two-position calibration manifold actuated by momentary pulses of control pressures. When placed in the calibration position, all sensors are connected to a common calibration pressure port. A series of accurately measured pressures can be applied through this port to characterize the sensors in place, at the conditions to which they are being subjected. Proper and periodic on-line calibration maintains static errors within +/-0.03% or better of the full scale (FS) pressure range. The calibration manifold automatically switches between two common reference pressures to ensure a stable reference is applied during calibration. The calibration manifold incorporates a visual position indicator and utilizes a dynamic O-Ring seal design for high reliability and easy servicing in the field.
Options include purge capability, isolated from the pressure sensors, to clear measurement lines of moisture and contamination, as well as to prevent closure of model pressure taps during application of pressure-sensitive paint. Other options include a nickel foil RTD output to monitor internal scanner temperature and Digital Temperature Compensation technology, now incorporating PSI’s Gen-2 amplifier technology.
purge Option
The Digital Temperature Compensation (DTC) option incorporates the latest PSI innovations in ESP technology including temperature signal outputs from each silicon pressure sensor, embedded EEPROM retaining factory calibration information including temperature characteristics, programmable dual gain amplifier with optional dual range calibration, calibration manifold position sensing, and internal sensor excitation regulation with sense output. DTC Series ESP scanners are supported by PSI’s DTC Initium System and the SDI configuration of the System 8400 to dramatically reduce calibration requirements. The ESP scanners are also supported by a variety of third party compatible systems.
Type | Analog |
Number of Channels | 32 |
Operational Mode | Differential |
Full-Scale | 1 psi |
2.5 psi | |
5 psi | |
10 psi | |
15 psi | |
30 psi | |
45 psi | |
50 psi | |
100 psi | |
Output | ESP |
ESP-DTC | |
Output Interface | Analog |
Precision | ±0.03 % FS |
±0.06 % FS | |
±0.15 % FS | |
Excitation | 5 V DC |
12 V DC | |
Technology | Silicon Piezoresistive |
- In-Model wind tunnel testing
- Aircraft Flight testing
- Automotive track testing
32HD | 32-Channel Pressure Scanner |
32HC-0200000000 | ESP-32HD-DTC Heater Chamber |
32HC-0220000000 | ESP-32HD DTC Heater Chamber; Gen 2; (NCNR) |
Specifications for 32HD-DTC & 64HD-DTC Miniature Scanners
Note - the performance stated is after 1 hour of power to the scanner and Reference Port at atmospheric pressure. The Heater Chambers for the DTC Scanners are only needed to maintain the temperature within the Operating Temperature Range.
NOTES:
1. Consult factory for pressure ranges not listed. For pressure ranges above 50psi, these configurations are not compatible with the Initium product but
are okay to use with other systems (8400, Optimus).
2. Maximum applied pressure without reducing sensor performance.
3. Static accuracy includes combined errors due to non-linearity, hysteresis, and non-repeatability.
4. Achievable accuracy after span calibration. Consult applicable datasheets for DTC Series ESP specifications when used with legacy PSI System 8400
SDI configuration, DTC Initium or Optimus Systems.
5. Typical performance using 2nd order equation of the form Px=C0+C1(Vx)+C2(Vx2).
6. Typical performance using 4th order equation of the form Px=C0+C1(Vx)+C2(Vx2)+C3(Vx3)+C2(Vx4).
7. Accuracy is expressed as a percent of the deranged full scale when dual range calibration is specified.
Ordering Information
Pressure Scanner:
PN: 32HD-AABBCCDDEE ESP-32HD Scanner, 32 channels
PN: 64HD-AABBCCDDEE ESP-64HD Scanner, 64 channels
Order Example: 32HD-0103021120 ESP-32HD Scanner, ±10"WC, .063", 90º, purge, Digital Temperature Compensation with single range calibration
ESP-32HD Miniature Analog Pressure Scanner Datasheet Datasheet |
- Differential or Vented Gauge
- Dry Noncorrosive Gases
- Proof Pressure 4 Times Rated Pressure (≤ 30 psi)
- Proof Pressure 1.5 Times Rated Pressure (> 30 psi)
- 0.040" or 0.063" OD Stainless Steel Tubulation
- Digital Temperature Compensation (Option)
Type | Analog |
Number of Channels | 64 |
Operational Mode | Differential |
Full-Scale | 1 psi, 2.5 psi, 5 psi, 10 psi, 15 psi, 30 psi, 45 psi, 50 psi, 100 psi |
Output | ESP, ESP-DTC |
Output Interface | Analog |
- 16 differential channels in a 2.75 × 0.65 × 0.19 in body
- Factory Digital Temperature Compensation (DTC)
- Static accuracy up to ± 0.05 % FS (ranges > 5 psi)
- Optional dual-range calibration for triple sensitivity
- Compatible with Optimus, Initium, FDS, 8400 & most DTC DAQs
- Low-voltage (± 5 V) supply option
- Direct-mount version for flush, low-tubing installations
Type | Analog |
Number of Channels | 16 |
Operational Mode | Differential |
Full-Scale | 1 psi, 2.5 psi, 5 psi, 10 psi, 15 psi, 30 psi, 45 psi |
Output | ESP-DTC |
Output Interface | Analog |