Optimus, Pressure Scanning Data Acquisition System


- Wind-tunnel Pressure Scanning System
- ESP Pressure Scanner Compatible
- Up to +/- 0.03% Accuracy
- Up to 2000 Hz/port EU Data Throughput
- DTC Digital Temperature Compensation Technology
- Miniature Scanner Digitizing Interface
- On Line Calibration
- On Line Verification
- 8400 software compatibility
- IEEE-1588 PTPv2 Timing Protocol
- Sensor excitation feedback
- Fiber optic communication
- High data rates
The Optimus Pressure Scanning System is a modular, high-accuracy data acquisition platform engineered specifically for aerodynamic and wind tunnel test environments. Designed to replace and expand upon legacy 8400 systems, Optimus integrates advanced pressure scanners—such as the ESP and MicroScanner series—with high-throughput digital interfaces and precise calibration hardware.
Pressure inputs are digitized locally within the Miniature Scanner Digitizing Interface (mSDI), then transmitted via fiber optic cable to the Optimus System Processor (OSP). The system supports up to 32 pressure scanners, or 2048 channels, and offers both binary and engineering unit data formats over Ethernet (TCP/UDP), with rates up to 1 Gbps.
Key components include:
-
OSP (System Processor): Rack-mount unit managing all data processing and external communications.
-
OFIU: Fiber optic interface units inside the OSP, enabling EMI-free connection to mSDI modules.
-
mSDI: Digitizes scanner inputs and distributes power. Each unit supports up to 8 scanners.
-
RPS (Remote Power Supply): Provides precise voltage regulation with excitation feedback.
-
PCU (Pressure Calibration Unit): Delivers high-accuracy calibration pressure across multiple ranges.
-
RP (Remote Processor): Interfaces PCU units and manages pneumatic backplanes.
System Diagram
Type | Analog |
Number of Scanners | 8 |
32 | |
Output Interface | Ethernet |
Sample Rate | 1200 Hz per channel |
Optimus OSP | Optimus System Processor for Windtunnel Pressure Scanning System |
OFIU | Optimus Fiber Optic Interface Unit provides communication between the OSP & mSDI scanner interface via multimode fiber cable |
MSDI | Miniature Scanner Digitizing Interface is the connection point for the cables leading to the pressure scanners |
RPS 8491 | Remote Power Supply provides regulated power to mSDI that distributes power to connected pressure scanners +12, -12, & +5VDC. |
PCU | Pressure Calibration Unit is a high accuracy servo-controlled pressure generation instrument |
RP | 8400 Remote Processor (RP) are rack mounted interfaces for PCU's & PSU's |
LS | Local Slave can be connected to the Remote Processor to expand the number of PCU’s |
OSCB | Optimus Scanner Cable, connects ESP pressure scanners to mSDI scanner interface |
OFCA | Optimus Fiber Optic Cable, provides data communication between mSDI scanner interface & OFIU |
8476 | Power Cable, 8476 provides regulated power to the mSDI from the RPS. |
OPCY | Y-Cable, connects between the 8476 cable and mSDI |
PBU | Pneumatic Blanking Unit, is needed to fill unused slots in the RP or LS |
MM | Mechanical Models, used during model build in place of scanners and mSDI |
Optimus-Pressure-Scanning-System_B-ATI Datasheet |
- Support for 8 DTC Series ESP Scanners, 32 or 64 port Scanners
- For up to 512 Pressure Channels
- ±0.05% FS Precision after Rezero
- 1000 Hz per Channel Throughput in Engineering Units for 128 ports
- Syncronized Data with Accelerometers, Wet Pressures, etc.
- Auto-negotiating 10/100 Base-T Ethernet Interface
- Compact Robust Package, with an optional Protective Enclosure
- Comprehensive Applications Utility Software
Type | Analog |
Number of Scanners | 8 |
Number of Channels | 512 |
Output Interface | Ethernet |
Sample Rate | 1200 Hz per channel |
Excitation | 18 - 36 V DC |