Loading...
A-Tech Instruments Ltd.
Français   |   Welcome. Please Register an Account or Login

2329, Fast Resistance Measurement

The device operates on the basis of proven 4-wire technology which corrects any incoming line or contact resistances. The measurement lines are checked by an integrated cable rupture monitor. Needless to say, the functions include...
2329, Fast Resistance Measurement
2329, Fast Resistance Measurement
Features:
  • Measurement range from 200 mΩ ... 200 kΩ
  • Resolution up to 10 μΩ
  • Measurement accuracy 0.03 % rdg.
  • Automatic measurement range selection
  • Up to 50 measurements and evaluations per second
  • Temperature compensation for all materials
  • Thermovoltage compensation
  • Dry-circuit measurement in accordance with DIN IEC 512
  • Data logger for 20 000 measurement values
  • RS232 and PLC interface standard (IEEE488, USB and Ethernet option)
Manufacturer:  Burster Praezisionsmesstechnik Gmbh
Quote
Datasheet
Description
Specifications
Applications
Models & Accessories
Downloads
Similar Products
2329 - Description
The device operates on the basis of proven 4-wire technology which corrects any incoming line or contact resistances. The measurement lines are checked by an integrated cable
rupture monitor.

Needless to say, the functions include temperature compensation for various test substances such as copper, brass, tungsten etc. Temperatures are measured by a Pt 100 sensor or infrared sensor (pyrometer) with analog outputs or any temperature transmitter e.g. thermocouple.

For measuring test objects with low inductances, a special measurement input protection was developed to prevent voltage peaks from damaging the device when test objects are disconnected.

If objects need to be tested for several different parameters by an automatic measurement system, you can store up to 32 device settings such as measurement ranges, limiting values, temperature coeffi cients etc. These settings are called up via a bit pattern (5 bits).

Naturally, all device settings can also be made via the RS232 (standard) or IEEE488/USB (optional) interface.

The integrated data logger can be used during serial measurements to store up to 20 000 measurement values which can be divided into 32 individual blocks. A digital fi lter is available for preselecting the measurement values to be stored.

The evaluation menu of the stochastic data logger displays the maximum, minimum and average values as well as the standard deviation.
2329 - Specifications
Range200 mΩ
2 Ω
20 Ω
200 Ω
2 kΩ
20 kΩ
200 kΩ
InterfaceRS-232
USB
Ethernet
Digital I/O
IEEE 488
PLC
Supply110 V AC
230 V AC
2329 - Applications
The RESISTOMAT® model 2329 is particularly suitable for quick measurements of low resistances in automated processes. Up to 50 measurements per second can easily be performed.

The device complies with the latest CE guidelines and is designed for laboratory use as well as heavy-duty industrial applications.

A 2-fold and 4-fold comparator with switching outputs has been included for classifi cation and selection; this feature is particularly useful for serial experiments.

One special application involves the measurement of contact resistances (dry-circuit measurement); the load voltage in this case is limited to 20 mV in order to prevent fritting (DIN IEC 512).

The RS232 (standard) and IEEE488 (option) computer interfaces allow fully automatic testing stations to be set up. The PLC interface allows easy integration into your production process control.

Typical applications are resistance and conductivity measurements
on:
- Fuses
- Airbag triggers
- Coils for the automobile industry and electrical engineering
- Plug-in contacts and switches
- Commutator welded joints
- Meter samples in cable manufacture
- Circuit-board conductors etc
2329 - Models & Accessories
Models
2329RESISTOMAT® with RS232 interface
2329-V001RESISTOMAT® with RS232 and IEEE488 interface
2329-V002RESISTOMAT® with RS232 interface and 6 channel switch system
2329-V801RESISTOMAT®; for the check of airbag igniters, measuring current < 10 mA, voltage load < 2 V
2329 - Downloads
2329 - Similar Products
  • Measuring ranges from 2 mΩ to 200 kΩ
  • Resolution up to 0.1 μΩ
  • Accuracy 0.03 % Rdg.
  • Autorange
  • Temperature compensation for all materials
  • Thermal e.m.f. compensation
  • Input voltage protection up to 400 Vrms
  • RS232 and PLC interface standard (USB and Ethernet option)
Range
2 mΩ,
20 mΩ,
200 mΩ,
2 Ω,
20 Ω,
200 Ω,
2 kΩ,
20 kΩ,
200 kΩ
Error
≤ 0.03 %
Interface
RS-232,
USB,
Ethernet,
Digital I/O,
PLC
Supply
110 V AC,
230 V AC
  • Automatic choice of measuring ranges from 200 ohms to 20 ohms
  • Resolution up to 1 ohm¸
  • Standard interfaces IEEE488, RS232, RS485, USB (option)
  • Checking of tolerances, classification with statistics
  • Measuring error ≤ 0.01 %
  • Future-orientated measuring method with thermal e.m.f. compensation.
  • High level of stability due to constant comparisons with internal reference values.
  • Current regulation results in voltage-free disconnection, calculation of cooling curves of coils.
Range
200 µΩ,
2 mΩ,
20 mΩ,
200 mΩ,
2 Ω,
20 Ω,
200 Ω,
2 kΩ,
20 kΩ
Error
≤ 0.01 %
Interface
RS-232,
RS-485,
USB,
Ethernet,
IEEE 488
Supply
110 V AC,
230 V AC
  • Measuring ranges of 20 mΩ… to 200 kΩ
  • Resolution up to 1 µΩ & accuracy 0.03 % of reading
  • High-speed measurements from 10 ms/measurement, including evaluation
  • Temperature compensation for all materials
  • Thermoelectric voltage compensation
  • Input protection up to 400 V
  • 32 adjustable measuring programs
Range
200 mΩ,
2 Ω,
200 Ω,
20 kΩ,
200 kΩ
Error
≤ 0.03 %
Interface
USB,
Ethernet
Supply
110 V AC,
230 V AC